9:58 am

Hello

Hello

9:59 am

I get back to you about the ORP-01-3-0098 probe that I bought

9:59 am

Here is the problem: I manage to calibrate it at 470mv with a buffer solution

10:01 am

then, with the 650mV buffer solution it gives approximately 620mv - not very well, but above all, it does not take good measurements in real conditions, ie with the pool water: gives 550mv instead of 690 for example: - (

First thing: if you connect the probe and you dip it in the 470 mV solution, can you read 470 mV on the device?

10:02 am

yes to the first thing - after calibrating the device, because it changes a bit from one probe to another

10:04 am

calibration with the device

So., It seems that the probe is measuring the value at 470 mv. In this case put the pool water in a glass and then look at the value measured by the device

10:05 am

it gives for example 550 instead of 690

How do you know it should be 550?

10:07 am

this should be 690 because I got another probe, which calibrates precisely on 470 and 650, and because it corresponds to the results of tests with strips

10:09 am

to see otherwise: we have 3 samples

10:09 am

buffer solutions 470 and 650, and swimming pool water

In do one thing: take our probe: put it in the solution at 470 mV, look at the value measure. then put the second probe and look at the value measure

10:10 am

ok i do that

Keep me posted .... Thank you

10:16 am

Ok, the probe is working perfectly

10:18 am

Thank you