 | May 10, 2022, 12:13 pm Hello I recalibrated the ACL probe Despite this, it indicates on the redox alarm unit - 49 flashing (the probe is 3 years old) Thank you for your help |
 | Olivier Michaud May 10, 2022, 12:16 pm |
 | Olivier Michaud May 10, 2022, 12:17 pm |
 | May 10, 2022, 12:25 pm |
 | Olivier Michaud May 10, 2022, 12:26 pm |
 | May 10, 2022, 12:31 pm Hello I recalibrated the ACL probe Despite this, it indicates on the redox alarm unit - 49 flashing (the probe is 3 years old) Thank you for your help |
 | Olivier Michaud May 10, 2022, 12:32 pm |
 | May 10, 2022, 12:33 pm Ok I will do this test and get back to you. Thanks |
 | Olivier Michaud May 10, 2022, 12:33 pm |
 | May 10, 2022, 12:45 pm I just did the test: In water 208 mv In water with salt Alr 48 mv but the value drops slowly |
 | Olivier Michaud May 10, 2022, 12:47 pm The value of 208 mV is much too low. Do you have a 700mV standard solution? |
 | Olivier Michaud May 10, 2022, 12:47 pm |
 | May 10, 2022, 12:52 pm Last season the calibration was From 720 Mv |
 | Olivier Michaud May 10, 2022, 12:53 pm I wanted to know what you read when you put the probe in a standard solution |
 | May 10, 2022, 12:56 pm |
 | Olivier Michaud May 10, 2022, 1:09 pm The standard value is 700mV? |
 | Olivier Michaud May 10, 2022, 1:10 pm If yes then indeed the probe is tired |