Your questions about pH and Redox probes for swimming pools ph-redox-pool.com |
|
 | bot: Question EN Jul 18, 2022, 9:03 am . If you have any questions, please let us know. |
 | Jul 18, 2022, 9:03 am |
 | Jul 18, 2022, 9:03 am I feel like my. ke probe no longer works... |
 | Olivier Michaud Jul 18, 2022, 9:03 am |
 | Olivier Michaud Jul 18, 2022, 9:03 am |
 | Jul 18, 2022, 9:05 am I do the calibration it works but I have the impression that it does not want to send chlorine |
 | Olivier Michaud Jul 18, 2022, 9:06 am What regulator do you have? |
 | Jul 18, 2022, 9:06 am Is it due to my vial which is no longer good |
 | Jul 18, 2022, 9:06 am |
 | Olivier Michaud Jul 18, 2022, 9:08 am If you are injecting chlorine then it should be a redox sensor. can you give me the astralpool model |
 | Jul 18, 2022, 9:09 am |
 | Olivier Michaud Jul 18, 2022, 9:09 am |
 | Jul 18, 2022, 9:10 am |
 | Olivier Michaud Jul 18, 2022, 9:10 am |
 | Jul 18, 2022, 9:11 am |
 | Olivier Michaud Jul 18, 2022, 9:11 am |
 | Jul 18, 2022, 9:11 am Currently it displays 730 and suddenly as it does not take liquid chlorine I compensate with chlorine tablet |
 | Jul 18, 2022, 9:12 am |
 | Olivier Michaud Jul 18, 2022, 9:12 am a teaspoon. This test will allow us to know if the probe is still working |
 | Jul 18, 2022, 9:12 am If it drops more than 100mv it is good, right? |
 | Olivier Michaud Jul 18, 2022, 9:13 am |
 | Jul 18, 2022, 9:14 am ok i will test it right away |
 | Jul 18, 2022, 9:16 am |
 | Jul 18, 2022, 9:16 am Do you have to leave it long |
 | Olivier Michaud Jul 18, 2022, 9:17 am |
 | Jul 18, 2022, 9:17 am I went down to 370... and the roller works fine |
 | Jul 18, 2022, 9:17 am |
 | Olivier Michaud Jul 18, 2022, 9:18 am Do you know the set point? |
 | Jul 18, 2022, 9:18 am |
 | Jul 18, 2022, 9:18 am Maybe this is the solution that is no longer valid? |
 | Olivier Michaud Jul 18, 2022, 9:19 am |
 | Jul 18, 2022, 9:19 am |
 | Olivier Michaud Jul 18, 2022, 9:19 am |
 | Jul 18, 2022, 9:19 am Yes it spun when I put it in the glass |
 | Olivier Michaud Jul 18, 2022, 9:19 am And if you put it in the standard solution, how much do you read? |
 | Jul 18, 2022, 9:21 am |
 | Olivier Michaud Jul 18, 2022, 9:22 am The solution is at 468 mV? |
 | Jul 18, 2022, 9:22 am so i think everything is ok |
 | Jul 18, 2022, 9:22 am |
 | Olivier Michaud Jul 18, 2022, 9:23 am Me too . The pump does not start because the measurement is 730 mV and your set point is less than 730 mV |
 | Jul 18, 2022, 9:23 am Ok do I put 4560 oub460 in deposit? |
 | Jul 18, 2022, 9:23 am |
 | Olivier Michaud Jul 18, 2022, 9:24 am |
 | Jul 18, 2022, 9:24 am |
 | Olivier Michaud Jul 18, 2022, 9:26 am 468 is the calibration value, not the set value. Do not add any more chlorine, if all goes well the reading should go down and when the value is lower than the setpoint then the pump will start |
 | Jul 18, 2022, 9:26 am I still feel like I haven't used chlorine all summer. |
 | Jul 18, 2022, 9:27 am Ok it works I can redo a calibration or for you it's good no need I put the probe back in its circuit? |
 | Olivier Michaud Jul 18, 2022, 9:28 am It's good you measure 460 in the 468 mV solution. It's OK . |
 | Jul 18, 2022, 9:28 am Ok thank you I put everything back in function then 😜 |
 | Olivier Michaud Jul 18, 2022, 9:28 am |
 | Jul 18, 2022, 9:29 am , ok it works thank you very much for the time spent troubleshooting have a good day thank you again |
 | Olivier Michaud Jul 18, 2022, 9:29 am |
 | Olivier Michaud Jul 18, 2022, 9:30 am You are welcome, do not hesitate to contact us! Good day |
|
|
| |
|